Skip to content

Add hangUpOnNextMessages: hang up on a message, before or after answering - #33

Open
ps2 wants to merge 1 commit into
mainfrom
drop-during-next-command
Open

ps2 wants to merge 1 commit into
mainfrom
drop-during-next-command

Conversation

@ps2

@ps2 ps2 commented Nov 10, 2026

Copy link
Copy Markdown

Adds hangUpOnNextMessages: on each of the next count messages, the pod hangs up, then takes the phone's reconnect as usual.

  • {"command": "hangUpOnNextMessages", "value": 1} hangs up when the message arrives, without answering, as a pod that drops the link during a command does.
  • With "afterAnswering": true it answers, waits for the phone's ACK, then hangs up, as a pod that drops the link soon after a command does.
  • On -transport impossible the phone gets CoreBluetooth's "peripheral disconnected" error (CBError 7), the way a pod's hang-up reaches a phone. On a Bluetooth adapter it's an ordinary disconnect from the pod.

The simulator could already leave a message unanswered with the link up (ignoreNextMessages), fail a connect during setup (failNextConnects), or exit (crashNextCommand). None of those is a pod-initiated drop that leaves the pod available, which is the case in the loopandlearn/OmnipodKit#209 review: a pod that hangs up during or after a command, with the phone in the background.

Testing

  • go test ./pkg/... passes. A new test checks that the hang-up carries CBErrorDomain 7 and that the next connect is accepted.
  • Used against Loop (next-dev with Log each connect's setup and every failed exchange and disconnect; don't mark unsent commands unacknowledged loopandlearn/OmnipodKit#209) in the iOS Simulator through ImpossiBLE, with the app in the background on a heartbeat probe:
    pod:  hanging up on message seq 4 without answering (hangUpOnNextMessages)
    pod:  ** disconnected by the pod
    app:  [disconnect] CBErrorDomain 7, 0.010s after connect, during session Get pod status, setup ok
    app:  [read] emptyValue at RTS after 5.084s, link already down; …
    app:  Unacknowledged message. seq:6, error = incorrectResponse
    
    and with afterAnswering, the status read completes and the drop follows it:
    pod:  hanging up after answering message seq 4 (hangUpOnNextMessages)
    app:  [disconnect] CBErrorDomain 7, 0.017s after connect, during session Get pod status, setup ok
    

…ring

A pod in the field sometimes drops the link during a command, or soon
after one. The simulator could leave a message unanswered with the link
up, fail a connect during setup, or crash, but not hang up and stay
available. With this command it hangs up, with CBErrorPeripheralDisconnected
on the ImpossiBLE transport, on each of the next count messages: when the
message arrives, or with afterAnswering once the phone has acknowledged
the answer. It accepts the phone's reconnect either way.
Sign up for free to join this conversation on GitHub. Already have an account? Sign in to comment